US-EPA Approved for PM2.5, PM10 and PM10-2.5
Sampling Method Flow Rate 16.67 LPM
PM2.5 Reference Method (RFPS-0717-245)
US-EPA-PM10 Reference Method (RFPS-0717-246)US-EPA PM10-2.5 FRM Reference Method (RFPS-0717-247)
The E-SEQ-FRM accommodates up to sixteen (16) 47-mm diameter filter samples and may be pre-programmed to a flexible, user-defined sampling schedule permitting more than 2 weeks of unattended daily operation.
The E-SEQ-FRM is modular in design, and may be easily disassembled and transported from one site to another. Its compact design and small footprint and allows it to be moved in and out of locations with challenging ingress/egress. No disassembled component: system box, stand, or inlet assembly weighs more than 45 lb.
The E-SEQ-FRM offers an easy, intuitive event manager that permits programming of up to 16 independent sampling events. Optional modem/communication package allows for remote access of stored data via RS232 interface. Data is also quickly and easily retrieved with a USB flash drive.
Applications:
• Regulatory criteria reference sampler networks: Federal, State, Local.
• Academic studies
• Comparative studies
SPECIFICATIONS:
Method
Multi-event, sequential sampling
of ambient air onto filter media
Regulatory
Designations
US-EPA
Korea Certification
(APSC1-2016-01)
Filter
Holders and Media
47 mm filter For US-EPA PM2.5
and PM10-2.5 sampling: Teflon 2µm pore size membrane filter (regulation
required). For US-EPA PM10 reference sampling Pallflex TX-40 quartz filter
and Teflon media.
Sample
flow
Active flow control for 16.67
LPM. Sampled volume reported under actual and standard conditions
Calibration
Temperature, barometric pressure
and multi-point flow.
Internal
Data Storage
More than 130 days of interval
data with 5 minute time resolution 132 filter records
Data
Input/Output
Keypad/display for data
retrieval and programming. RS232, USB serial and USB flash drive interface
for data download
Operational
Temperature
-40°C to +50°C
Power
Requirements Physical
100-240 VAC 50/60 Hz 1.4 amps
Dimensions
(inches) without
System/Pump: 14.4 (w) x 14 (d) x
27 (h) Stand (at base)
PM10
inlet attached
24.5 (w) x 23.5 (d) x 29.6 (h)
Total (at base) 23.5 (w) x 24.5 (d) x 56.6 (h)
Weight:
System/Pump 45 lb. Stand 10.5
lb, PM10 Inlet 5.3 lb
Sequential Reference Gravimetric Particulate Sampler E-SEQ-FRM by Metone
- Brand: Met One Instruments Inc.
- Product Code: E-SEQ-FRM
- Availability: Pre-Order
Related Products
Automatic Continuous Real Time Particulate Monitor BAM 1020 by Metone
US-EPA Approved FEM, TUV Approved for PM10, PM2.5, PM 10-2.5 & TSP Monitoring Flow Ra..
Portable Automatic Continuous Real Time Particulate Monitor BAM 1022 by Metone
US EPA Approved for PM2.5 Class III FEM Ref (EQPM-1013-209) Flow Rate 16.7 l/m BAM 1..
Portable Dust Monitor for PM 2.5, PM 10, TSP by Metone
The BT‐645 Portable Dust Monitor has been designed to continuously measure and record particul..
Portable Continuous Real Time E BAM Monitor by Metone
Accuracy and precision consistent with U.S.-EPA requirements for Class III PM2.5 and PM10 meas..
SASS/SUPER SASS Particulate Air Sampler by Metone
US-EPA Certified, 5 & 8 Channel Gravimetric , Flow Rate 6.7 liter/min Two models prov..
Portable Automatic Continuous Real Time Particulate Monitor E-BAM Plus by Metone
US EPA Approved FEM Method for PM10 The Met One Instruments, Inc. E-BAM Plus Portable Bet..
Tags: Sequential Reference Gravimetric Particulate Sampler E-SEQ-FRM by Metone